Lunge窒素計法における補正について
スポンサーリンク
概要
- 論文の詳細を見る
The amounts which are necessary to be corrected in the determination of nitrogen by the Lunge method have been investigated. The loss of nitrogen oxide by its solubility in 8797% sulfuric acid was determined by the titration method and by the colorimetric method. The effect of temperature on the solubility was rather small. In the determination of nitrogen in nitric acid, a correction for dissolved nitrogen oxide gave a result in agreement with that determined by acidimetric method. In this case, the formation of nitrous oxide by side reaction is compensated for by the sulfuric acid remaining adhered to the wall of the burette. However, a nitrocellulose sample yielded carbon monoxide and carbon dioxide by side reaction and gave +0.2% error with the correction for solubility of nitrogen oxide and -0.3% error without the correction. Hence, a proposed correction formula for the amount of gas in the Lunge nitrometer is given.
- 社団法人 日本分析化学会の論文
著者
関連論文
- ガスクロマトグラフィーによる直鎖型アルキルベンゼンスルホン酸塩の定量
- フェロイン・ブロモフェノールブルーイオン会合抽出を利用するリンの間接高感度吸光光度定量
- ニトラニル酸バリウムを用いる硫酸イオンの間接吸光光度定量
- p-アミノアゾベンゼンを用いる亜硝酸イオンの吸光光度定量(環境汚染物質の分析化学)
- クリスタルバイオレット-クロラニル酸抽出法による硫酸イオンの間接吸光光度定量
- クロラニル酸バリウムを用いる硫酸イオンの高感度吸光光度定量
- メスバウアースペクトル法及びX線回折法による尿素を含む溶液から均質的に沈殿した鉄化合物の化学状態の分析
- 高速液体クロマトグラフィーと赤外粉体反射スペクトロメトリーとの結合測定法
- 工業廃水中の有機水銀化合物抽出法の改良
- 薄層クロマトグラフ法による工業廃水中の有機水銀化合物の分離
- 硫酸第一鉄滴定による硝酸および硝酸繊維素中の窒素迅速定量法 : 硝酸および硝酸繊維素中の窒素定量法研究(第6報)
- 吸光光度法による硝酸繊維素中の窒素の定量
- 酢酸繊維素の酢化度測定の迅速化
- 硝酸および硝酸繊維素中の窒素の容量分析 : 硝酸および硝酸繊維素中の窒素定量法研究(第1報)
- Lunge窒素計法における発生酸化窒素の混入成分について
- Lunge窒素計法における発生気体の組成について
- アニリンを用いる混酸中の硫酸迅速定量法
- 工業ガス分析 (進歩総説)
- チオシアン酸スカンジウムのエーテル抽出