414 遠心血液ポンプの動圧軸受形状がインペララジアル安定性と溶血に及ぼす影響(OS5-(4)オーガナイズドセッション《ライフサポートにおける工学技術》)
スポンサーリンク
概要
- 論文の詳細を見る
We have been developing a centrifugal blood pump with hydrodynamic bearings. The purpose of this study is to clarify experimentally the effect of the configuration of a hydrodynamic journal bearing to impeller radial stability and hemolysis for centrifugal blood pumps with hydrodynamic bearings for two weeks circulatory assist in order to establish a hydrodynamic journal bearing design method. We compared two depths of herring-bone groove on the journal bearing as 50 μm and 0 μm (no groove). Impeller trajectory was measured using two laser displacement sensors. Next, an in vitro hemolysis test for 4 hours was conducted with bovine blood added with heparin at a flow rate of 4 L/min and a pressure head of 200 mmHg. The experiments showed that the groove on the journal bearing will be useful for hemolysis reduction resulted by the impeller radial stability in a centrifugal blood pump.
- 2011-08-26
著者
-
西田 正浩
産総研
-
山根 隆志
産総研
-
丸山 修
産総研
-
小阪 亮
産総研
-
川口 靖夫
東理大
-
Taenaka Yoshiyuki
Department Of Artificial Organs Research Institute National Cardiovascular Center
-
山根 隆志
総合機械:産総研
-
Yamane Toshikazu
Department Of Mechanical Systems Engineering Graduate School Of Engineering Toyama University:(prese
-
安井 和哉
東理大院
関連論文
- 133 MERAモノピボット遠心ポンプの血液適合性に関わる内部流れの数値解析手法に対する検討(OS1-5:循環器系・血流の計測とメカニクス(5),OS1:循環器系・血流の計測とメカニクス)
- 135 閉回路を利用した遠心血液ポンプの抗血栓性評価(OS1-5:循環器系・血流の計測とメカニクス(5),OS1:循環器系・血流の計測とメカニクス)
- 136 補助循環用モノピボット遠心ポンプの抗血栓性と耐久性(OS1-5:循環器系・血流の計測とメカニクス(5),OS1:循環器系・血流の計測とメカニクス)
- 補助循環用遠心ポンプの壁近傍せん断応力の測定
- 303 MERAモノピボット遠心ポンプの流れ解析(OS2-1:循環系のバイオメカニクス,オーガナイズドセッション2:循環系のバイオメカニクス)
- B101 臨床応用をめざすモノピボット遠心血液ポンプの血液適合性評価(B1-1 人工臓器)
- 補助循環用モノピボット遠心ポンプの抗血栓性評価
- MERAモノピボット遠心血液ポンプの流れの可視化解析
- 527 簡便なin vitro試験法による遠心血液ポンプの抗血栓性評価(GS-12 循環器工学,一般セッション,学術講演)
- 526 In vitro抗血栓性試験法確立のための血液凝固能制御方法の検討(GS-12 循環器工学,一般セッション,学術講演)