医用直線加速装置の高エネルギーX線による放射化の解析 : 短半減期放射化核種の推定
スポンサーリンク
概要
- 論文の詳細を見る
In medical linear accelerators, radioactivation is induced on the target and neighborhood parts by photoneutrons accompanying a photo-nuclear reaction and leading to higher acceleration energy. We measured the residual radiation from the radioactivated materials according to the time, and tried to identify radioactivated nuclides and their relative quantities by means of measurement results. It was presumed that the main source of residual radiations was the Target, Flattening filter and Primary collimator in the linac head. Among those materials (copper, tungsten), we calculated decrement curves of residual radiations from radioactivated nuclides generated with photo-nuclear reaction or thermal neutron capture reaction by various ratios, and we investigated the ratio that best fit the measured data. Consequently, it was presumed that 66Cu generated with thermal neutron capture reaction contributed the most to residual radiation, followed by 62Cu generated with photo-nuclear reaction contributed. It is important to understand various characteristics of these nuclides and to undertake management of the device.
- 公益社団法人 日本放射線技術学会の論文
- 2010-06-20
著者
-
川野 誠
名古屋市立大学病院中央放射線部
-
加藤 秀起
藤田保健衛生大学医療科学部放射線学科
-
鈴木 友輔
藤田保健衛生大学大学院保健学研究科
-
青山 貴洋
愛知医科大学病院中央放射線部
-
高橋 礼紀
新潟市民病院放射線技術科
-
福間 宙志
名古屋市立大学病院中央放射線部
-
磯山 茂
名古屋市立大学病院中央放射線部
-
福間 宙志
名古屋市立大学病院・中央放射線部
-
加藤 秀起
名古屋市立大学医学部附属病院 中央放射線部
-
川野 誠
金沢大学 大学院医学系研究科保健学専攻
-
川野 誠
名古屋市立大学大学院システム自然科学研究科
-
加藤 秀起
藤田保健衛生大 医療科学
関連論文
- 角度計測誤差がpresampled MTFへ及ぼす影響の検証と角度計測の最適手法の提案
- 医用直線加速装置の高エネルギーX線による放射化の解析–短半減期放射化核種の推定-
- 新しいIVR線量分布作成方法の開発
- CRポータルイメージにおける周波数処理の影響
- 診断領域における散乱 X 線のスペクトル的解析
- 高エネルギー X 線 Narrow beam の物理特性
- 14. 微分 1 次線および微分散乱線々量を用いた 3 次元線量計算法(放射線治療, 中部部会)
- 13. 治療計画用コンピュータへの人体横断面入力における精度について(放射線治療, 中部部会)
- Monte Carlo 法を用いた体内散乱線の解析 : 斜入射照射における散乱線々量の変化
- 診断領域 X 線に対する 1cm 線量当量換算係数