Various conditions in the determination of lead by atomic absorption spectroscopy were examined. Lead was extracted as iodide from phosphoric acid solution with MIBK with the extraction rate over 99%. The working conditions for MIBK extracts were: the wavelength 2833 Å, the current of hollow cathods lamp for Pb 9 mA, the slit width 0.18 mm, air pressure 1.8 kg/cm<SUP>2</SUP> (12<I>l</I>/min) and acetylene pressure 0.5 kg/cm<SUP>2</SUP> (2<I>l</I>/min). In the aqueous solution, the optimum air pressure was 1.8 kg/cm<SUP>2</SUP> (13<I>l</I>/min) and acetylene pressure 0.5 kg/cm<SUP>2</SUP> (3.5<I>l</I>/min).<BR>By the extraction method, the sensitivity was increased five-fold compared to that by the aqueous medium. The method was applied to the determination of lead in cast iron, free cutting aluminum alloy and zinc die cast with satisfactory results.