Diffusion-weighted Imaging of the Liver : Technical Challenges and Prospects for the Future
スポンサーリンク
概要
- 論文の詳細を見る
Diffusion-weighted imaging (DWI) has recently been attempted in the abdominal region. We review diffusion-weighted images of the liver, especially from the technical point of view. We discuss selection of pulse sequence parameters, effects of anti-breathing motion technique, tips for measuring apparent diffusion coefficient (ADC), and utility of superparamagnetic iron oxide (SPIO), showing clinical cases, including those at 3T. Our current trial of new pulse sequencing, such as SPIO-mediated breath-holding black-blood fluid-attenuated inversion recovery (BH-BB-FLAIR), imaging is shown. Some prospects for the future in DWI of the liver are also stated.
- 日本磁気共鳴医学会の論文
- 2005-12-31
著者
-
石口 恒男
名古屋大学 放射線
-
深津 博
名古屋大学医学部放射線医学教室
-
深津 博
名古屋大学放射線科
-
深津 博
国立療養所鈴鹿病院 神経内科
-
深津 博
名古屋大学医学部放射線科
-
NAGANAWA Shinji
Department of Radiology, Nagoya University Graduate School of Medicine
-
KAWAI Hisashi
Department of Radiology, Nagoya University Graduate School of Medicine
-
深津 博
名古屋大学医学部附属病院 放射線部
-
深津 博
名古屋大学 医学部放射線科
-
Kawai Hisashi
Department Of Radiology Nagoya University Graduate School Of Medicine
-
SAKURAI Yasuo
Department of Radiology, Teine Keijinkai Hospital
-
ISHIGAKI Takeo
Department of Radiology, Nagoya University School of Medicine
-
FUKATSU Hiroshi
Department of Radiology, Nagoya University Hospital
-
Fukatsu Hiroshi
Department Of Radiology Nagoya University Graduate School Of Medicine
-
石口 恒男
名古屋大学放射線科
-
Mimura Takeo
Department Of Radiology First Kamiida General Hospital
-
Naganawa Shinji
Department Of Radiology Nagoya University Graduate School Of Medicine
-
Sakurai Yasuo
Department Of Radiology Nagoya University Graduate School Of Medicine
-
Ishigaki Takeo
Department Of Radiology Nagoya City University Medical School
-
Ishigaki Takeo
Dept. Radiol. Fac. Med. Nagoya Univ.
-
AOKI Ikuo
Toshiba Medical Systems
-
MIURA Shunichi
Department of Radiology, Toyohashi Municipal Hospital
-
KANAZAWA Hitoshi
Toshiba Medical Systems
-
Miura Shunichi
Department Of Radiology Toyohashi Municipal Hospital
-
Fukatsu Hiroshi
名古屋大学医学部附属病院 放射線部
-
Sakurai Yasuo
Department Of Radiology Asahikawa Kosei General Hospital
関連論文
- 病診連携における医療コンシェルジュ導入による外来診療の改善効果と問題点
- 医療コンシェルジュ導入による外来診療の改善効果と問題点 : 外来待ち時間の劇的短縮と患者満足度向上を目指して
- IVRにおける患者の被ばく線量測定に関する基礎的検討(第31回秋季学術大会放射線防護管理関連演題発表後抄録)
- IVR実施4施設における術者の被ばく線量(第31回秋季学術大会放射線防護管理関連演題発表後抄録)
- IVR実施4施設における患者の被ばく線量(第31回秋季学術大会放射線防護管理関連演題発表後抄録)
- 肝細胞癌の動脈塞栓療法における患者と術者の被曝測定
- OP-1-105 MD-CTによる肝門部胆管癌における胆管内進展範囲の術前診断の検討(胆道3)
- Self-expandable metallic stent 留置拡張術を行った悪性腫瘍による気管・気管支狭窄の 3 例
- 3. 気管・気管支の悪性狭窄にたいする self-expandable metallic stent 留置拡張術(1) 気管・気管支狭窄, 第 6 回日本気管支学会中部支部会(第 38 回東海気管支鏡研究会))
- 遠隔医療・テレラジオロジー
- 低磁場MRIを用いた睡眠時無呼吸時の上気道と声帯の変化 : 多系統萎縮症と肥満による閉塞性睡眠時無呼吸症との対比
- O2-54 潜因性West症候群におけるDiffusion Tensor画像とFDG-PET所見の比較(画像3,一般演題(口演),第42回日本てんかん学会)
- V-59 大静脈バルン閉塞を用いた大動脈内Transluminally Placed Endovascular Graft留置法
- 78 IVR実施4施設における術者の被曝線量(放射線管理 IVR被ばく)
- 77 IVR実施4施設における患者の被曝線量(放射線管理 IVR被ばく)
- O-589 遠位弓部大動脈瘤に対するEndovascular Graft移植術の臨床検討
- 83) 直視下胸部大動脈ステントグラフト留置術における脊髄障害予防の工夫
- 正中神経電気刺激における第一次体性感覚野における脳活動について : fMRIを用いた検討
- 画像フィルターの 3D time-of-flight MR angiography の画質に及ぼす影響
- Single shot gradient echo EPIを用いた正常上腹部臓器のperfusion dynamic scanの検討
- 大動脈疾患に対するTransluminally Placed Endovascular Graftの治療成績
- MRIにて被殻病変を認め, 神経病理学的に確認した, 家族性パーキンソニズムの一家系
- Leigh脳症の1剖検例 : MRI拡散強調画像高信号の病理学的背景
- 大血管に対する Transluminally Placed Endovascular Graft 内挿術の中期成績
- 3. 術前・術中検査 ( 大動脈瘤に対するステント人工血管内挿術)
- 3D-Fast Spin Echo法を用いた内耳器官微細構造の観察
- PP189 狭心症を有する遠位弓部大動脈瘤に対するopen surgeryによるstent grafting と冠動脈バイパス術
- 小児CTガイドライン : 被ばく低減のために(ガイドライン)
- 画像診断レポートの新しい展開:Hyperlink report レポート上の語句とDICOM viewer上の画像のhyperlinkを簡単に作成でき、情報の迅速かつ確実・安全な伝達を可能とするレポートシステムの開発と評価 (特集2 PACSと電子カルテ)
- 低磁場MR fluoroscopyによる睡眠時無呼吸症候群の評価
- 選択的外頚動脈撮影法とその臨床的応用
- 離脱式バルーンカテーテルによる外傷性内頸動脈海綿静脈洞瘻の治療
- 日本医学放射線学会ガイダンスレベルの導入に必要なX線撮影の患者線量の評価法の研究-TLDによる入射表面線量の直接測定による調査-放射線防護委員会報告
- 医療コンシェルジュの導入による地域医療連携の効率化
- 1.腎疾患の画像診断腫瘍性疾患を中心に
- Interventional MRIの現状と将来(MRの治療応用と機器)
- 3D imaging の現状と将来
- 9. 肺癌症例における肺シンチグラフィによる考察 : 第32回日本肺癌学会中部支部会
- modern MRIにおけるartifact,peudolesion,pitfall--日常の読影における問題点と解決法
- 肝臓疾患のCT, USおよびMRIによる診断能の客観的評価に関する研究
- MRI
- フェリデックス投与後のT1強調像が評価に有用であった2例 (肝腫瘍性病変におけるMRI診断--フェリデックスMRI症例)
- 3D-fast spin echo法を用いた小脳橋角部のMR cisternography:280例の検討
- MRIが病変部の確定に有用であった視神経炎の一例
- 大学病院としての取り組み (特集 PACS 2008)
- Neural networkを用いた医用画像圧縮技術の開発と評価(第1報)Lossless圧縮における評価 (特集 PACSはここまで進化した)
- Hyperlink reportとORION (特集 PACSはここまで進化した)
- 3 Tesla MRIによる前立腺癌の診断--拡散係数画像の有用性 (Routine Clinical MRI 2006 Book)
- 検査増や画像の膨大化などによる放射線科医の労働強化 (総特集 今、放射線診療が抱える課題と展望)
- MRIおよびCTによる頭蓋内疾患診断の客観的評価に関する研究
- 診療放射線技術学科の臨床実習指導方法改善に関する調査研究報告
- 9. メタリック・ステント留置後の気管再狭窄にレーザー治療が奏効した肺癌の 1 例(第 16 回 日本気管支学会中部支部会)
- Influence of liquid crystal display monitors on observer performance for detection of diffuse pulmonary disease on chest radiographs
- Cerebral glucose metabolism change in patients with complex regional pain syndrome : a PET study
- Intra- and Interobserver Agreement and Performance Score of Breast Phantom Image Interpretation: Influence of Ambient Room Lighting Levels
- Application of Subsecond Rotation Scan to Helical CT for Lung Cancer Screening
- Evaluation of New Image Processing Conditions for Digital Mammograms from FUJI Computed Radiography
- Influence of Monitor Luminance and Room Illumination on Soft-Copy Reading Evaluation with Electronically Generated Contrast-Detail Phantom: Comparison of Cathode-Ray Tube Monitor with Liquid Crystal Display
- Three-phase CT Examination of the Pancreatobiliary Region Using Multislice CT with 1-mm Collimation
- Multiphase Contrast-enhanced CT of the Liver with a Multislice CT Scanner : Effects of Iodine Concentration and Delivery Rate
- Multi-slice CT of Thyroid Nodules : Comparison with Ultrasonography
- Evaluation of Abdominal Aortic Aneurysm for Endovascular Stent-grafting with Volume-rendered CT Images of Vessel Lumen and Thrombus
- Usefulness of a Method for Changing Tube Current during Helical Scanning in Multislice CT
- Statistical method in a comparative study in which the standard treatment is superior to others
- Evaluation of the space collaboration system : its history, image quality and effectiveness for joint case conference
- Screening Helical CT for Mass Screening of Lung Cancer: Application of Low-Dose and Single-Breath-Hold Scanning
- Digital Radiology and PACS
- マルチスライスCT透視を用いた経皮的生検とIVR (特集 IVRの進歩)
- 高速螺旋CTによる肺癌検診のための基礎的検討-第一報 ファントムによる被曝線量と画質の評価-
- 放射線防護の「最適化」とガイダンスレベル (特集 放射線と健康)
- Usefulness of FDG PET for diagnosis and radiotherapy of the patient with malignant lymphoma involving bone marrow
- An abnormal accumulation of fluorine-18-FDG PET in cytomegalovirus enteritis-A case report
- New Insight into the Analysis of 6[^F]fluoro-L-DOPA PET Dynamic Data in Brain Tissue without an Irreversible Compartment : Comparative Study of the Patlak and Logan Analyses
- The Role of Contrast-Enhanced MR Mammography for Determining Candidates for Breast Conservation Surgery
- A Method for Measuring the Dose Distribution of the Radiotherapy Domain using the Computed Radiography System
- Pancreatic Ductal Adenocarcinoma Showing Iso-attenuation in Early-phase Contrast-enhanced CT : Comparison with Histopathological Findings
- High- and moderately high-methionine uptake demonstrated by PET in a patient with a subacute cerebral infarction
- Interventional Radiologyの現況と将来展望
- Interactive Fusion of Three-dimensional Images of Upper Abdominal CT and FDG PET with No Body Surface Markers
- 初回生検陰性例におけるMRSpectroscopyを併用した前立腺再生検の検討
- The Effects of Static Magnetic Fields and X-rays on Instability of Microsatellite Repetitive Sequences
- Contrast-enhanced MR Imaging of Metastatic Brain Tumor at 3 Tesla : Utility of T_1-weighted SPACE Compared with 2D Spin Echo and 3D Gradient Echo Sequence
- Prompt Contrast Enhancement of Cerebrospinal Fluid Space in the Fundus of the Internal Auditory Canal : Observations in Patients with Meningeal Diseases on 3D-FLAIR Images at 3 Tesla
- High-speed Imaging at 3 Tesla : A Technical and Clinical Review with an Emphasis on Whole-brain 3D Imaging
- 経カテーテル的塞栓術(transcatheter embolization)により治療し得た先天性腎動静脈瘻の1例 : 第38回中部総会
- 脳腫瘍の放射線治療効果判定におけるCT scanの有用性
- Aortoiliac Stenooculusive Disease and Aneurysms: Screening with Non-contrast Enhanced Two-dimensional Cardiac Gated Cine Phase Contrast MR Angiography with Multiple Velocity Encoded Values and Cardiac Gated Two-dimensional Time-of-flight MR Angiography
- 12. メタリックステントを用いた気管・気管支狭窄の治療(第 10 回 日本気管支学会中部支部会)
- Stent-graft Placement for Pseudoaneurysm of the Aorta
- 腫瘤形成型膵炎のCT像の検討-病理組織所見との対比-
- Diffusion-weighted Imaging of the Liver : Technical Challenges and Prospects for the Future
- Clinical Evaluation of Three-dimensional MR-Cholangiopancreatography using Three-dimensional Fourier Transform Fast Asymmetric Spin Echo Method(3DFT=FASE): Usefulness of Observation by Multi-planar Reconstruction
- Contrast-enhanced MR Angiography of the Carotid Artery using 3D Time-resolved Imaging of Contrast Kinetics: Comparison with Real-time Fluoroscopic Triggered 3D-elliptical Centric View Ordering
- Imaging Endolymphatic Hydrops at 3 Tesla Using 3D-FLAIR with Intratympanic Gd-DTPA Administration
- Arachnoid Granulations on High-resolution MR Images and Diffusion-weighted MR Images:Normal Appearance and Frequency
- Water fat separation using the single acquisition "sandwich" type 3-point Dixon method to optimize knee joint scans
- Investigation on Invasive or Metastatic Potentials of Human Cancer Cells as a predictive assay for long term survival after radiotherapy (RT).
- High-resolution MR Images of Inner Ear Internal Anatomy using a Local Gradient Coil at 1.5 Tesla:Correlation with Histological Specimen
- Elongation of survival period of KSN mice bearing metastatic human uterine cervical cancer by TJ-48(Juzen-Taiho-to), a Japanese herbal. after radiotherapy.
- 39. 巨大脳動脈瘤および脳動静脈奇形に対するintraVaSCUIarrouteによる治療