A sensitive method for the preconcentration and determination of trace amounts of Co, Cu and Zn by energy-dispersive X-ray fluorescence spectrometry (EDXRF) has been developed. The method is based on the fact that 4-(2-pyridylazo)-resorcinol (PAR) loaded Dowex anion-exchange resin (PAR-resin) can effectively adsorb Co, Cu and Zn at pH 9.0 to form PAR-metal complexes. The detection limits for Co, Cu and Zn were 1.53, 0.31 and 0.21 ppb, respectively. The precisions for five replicate measurements of the three metals were 3.4, 2.7 and 2.1% RSD, and the calibration curves were linear up to 75 µg with correlation coefficients of 0.9975, 0.9980 and 0.9985, respectively. The method was successfully applied for the simultaneous determination of Co, Cu and Zn in seawater samples at ppb levels.