A Novel Method for the Assay of .ALPHA.-Glucosidase Inhibitory Activity Using a Multi-channel Oxygen Sensor.
スポンサーリンク
概要
- 論文の詳細を見る
The quantification of glucose by using a multi-channel dissolved oxygen (DO) meter (DOX96) with immobilized glucose oxidase (GOD) and mutarotase (MUT) was performed. An evaluation of the inhibitory activities for α-glucosidase (AGH) by modifying our batch-type pseudo-in vivo assay system [Oki et al.; Biol. Pharm. Bull., 2000, 232, 1084] was also performed using a DOX96. When 45 U/well GOD and 18.75 U/well MUT were immobilized on the surface of a gelatin membrane on the electrodes, the response shown by the decrease percent of DO (%) obtained with 8 electrode wells in the same row was linear with the glucose concentration up to 3.3 mM and a correlation coefficient larger than 0.9. To estimate the AGH inhibitory activity, AGH-immobilized Sepharose supports in the well of a silent screen plate were used. The IC50 values of acarbose and 1-deoxynojirimycin, a medicinal inhibitor for diabetes, were 0.70 ± 0.08 µM and 0.40 ± 0.13 µM, respectively, and coincided well with those by a pseudo-in vivo assay.