Implementation of a Generalized Standard Addition Method in a Flow Injection System Using Merging-Zones and Gradient Exploitation.
スポンサーリンク
概要
- 論文の詳細を見る
A novel strategy for implementing a generalized standard addition method (GSAM) in a simple two-channel flow injection (FI) system and for correcting matrix effects and spectral interferences in multicomponent analysis is described. The FI-GSAM system exploits the concentration gradients of the sample and the standard solution, requires only one standard solution per analyte and provides several addition levels using a single simultaneous injection of the sample and of each standard solution. Because the matrix effects change at every addition level, a novel mathematical model for this FIGSAM has been derived. The method was critically evaluated for simultaneous K, Ca and Na determinations by flame photometry in synthetic samples with different ethanol or glycerol contents, added to the samples to promote matrix effects. The results show good agreement with the expected values, with the relative error and the overall relative standard deviation usually being <5% (n=5) for 2.5 - 5.0 Na, 20.0 K and 1000 mg l-1 Ca. The elapsed time for the simultaneous determination of three analytes was about 2 min, consuming 400 μl of the sample and 100 μl of each standard solution.
- 社団法人日本分析化学会の論文
社団法人日本分析化学会 | 論文
- 微生物学的試験法による畜産物中に残留する抗菌性物質の高感度測定法(安心と安全に役立つ分析化学)
- 液体クロマトグラフィー/質量分析法による缶飲料中のビスフェノールAの定量 (環境ホルモンと分析化学)
- 高速液体クロマトグラフィーによるL-シトルリン測定を指標としたマクロファージの一酸化窒素産生の評価 ( バイオサイエンスと分析化学)
- 小型定電流電解電量/光度滴定装置の開発と食用油の過酸化物価定量への応用