フィルター上に捕集した大気粉じんのひょう量 : 湿度の影響とその対策
スポンサーリンク
概要
- 論文の詳細を見る
After airborne particulates are collected with a high volume sampler or a low volume sampler, it is prescribed that those samples are weighed in the atmosphere at 20 ℃ and 50 % relative humidity. Therefore, an experimental apparatus for the easy control of humidity was developed. From of the results of weighing of blank filters under the various values of humidity, it was proved that filters of cellulose ester had the properties of larger moisture-adsorption than those of glass fiber. The weighing of airborne particulates collected on filters is difficult because of moistureadsorption of those. But the samples were able to be easily and accurately weighed by being tightly covered with aluminum leaves in the box for conditioning of the samples before weighing of those. The results of weighing with this method were independent of various values of humidity in the room where those samples were weighed.
- 社団法人日本分析化学会の論文
- 1983-12-05
著者
-
西川 雅高
(社)日本分析化学会金属成分分析用土壌標準物質作製小委員会:環境庁国立環境研究所:(現)独立行政法人国立環境研究所化学環境研究領域
-
西川 雅高
環境庁国立公害研究所
-
溝口 次夫
環境庁国立公害研究所
-
宮坂 恵子
国立公害研究所計測技術部
-
宮坂 恵子
環境庁国立公害研究所
関連論文
- 金属成分分析用土壌標準物質の開発
- アンダーセンサンプラー用各種捕集材中の不純物量(超微量分析のための前処理と予備濃縮)
- 石炭フライアッシュの酸処理による元素抽出と表面組織の変化(石炭利用と分析化学)
- フィルター上に捕集した大気粉じんのひょう量 : 湿度の影響とその対策
- 蛍光X線分析法による大気浮遊粒子状物質中の金属の定量(環境汚染物質の分析化学)
- 中性ヨウ化カリウム法によるオキシダント測定法の温度効果及び濃度効果
- 紫外線蛍光法による二酸化硫黄の自動測定における励起光量の補償
- 誘導結合プラズマ発光分析法のための降水中の微量元素の蒸発濃縮法