逆相クロマトグラフィー-紫外吸光光度法によるレニウムの定量
スポンサーリンク
概要
- 論文の詳細を見る
Ultraviolet spectrophotometric determination of perrhenate ion is the simplest method for rhenium, but the method is subjected to interference from many ions which exhibit their absorption in the ultraviolet. By a reversed-phase column chromatography with tributyl phosphate-Teflon, perrhenate ion is separated form these ions, particularly molybdenum, and the method can be applied satisfctorily. The column with 6 mm in inside diameter is packed with 1 g of Teflon supports impregnated with 1 ml of tributyl phosphatc and equilibrated with 10 ml of 1 M sulfric acid solution. Acid concentration of the sample solution is adjusted about 1 M with sulfric acid and the solution is passed through the column. Molybdate and other metal ions are removed with 20 ml of 1 M sulfric acid solution, and the perrhenate ion is eluted with 25 ml of distilled water. The absorbance of the effluent is directly measured at 227 nm by means of flow cell (10 mm). The method is applicable to more than 20 μg of rhenium in molybdenite concentrates and its flue dust. When. the effluent is diluted to 25 ml in a volumetric flask and measured the absorbance with 10 mm cell, the method is applicable to more than 60μg of rhenium.
- 1975-09-10
著者
関連論文
- メチルバイオレットによるレニウムの抽出吸光光度定量法
- ネオクプロイン-銅(I)錯陽イオンを用いたレニウムの溶媒抽出吸光光度定量法と間接原子吸光定量法
- 逆相クロマトグラフィー-紫外吸光光度法によるレニウムの定量
- リン酸トリブチル逆相クロマトグラフィーによるレニウムとモリブデンの分離
- レニウム・タングステン合金中のレニウムの定量
- チオ尿素によるレニウムの吸光光度定量法
- トリフェニルメタン色素によるレニウムの抽出吸光光度定量法
- 鉄鋼および鋳鉄中の亜鉛の定量