波長選択性をもつInSb検知器を用いた遠赤外回折格子分光装置の試作
スポンサーリンク
概要
- 論文の詳細を見る
An n-type InSb detector in intense magnetic fields has very sharp spectral responsivity due to the cyclotron resonance absorption, so that the wavelength giving a maximum response is tunable with the applied magnetic field.A far-infrared grating spectrometer of Czerny-Turner type, of which f-number is 3, using this tunable InSb detector has been designed and constructed in the frequency range from 33 to 333 cm<SUP>-1</SUP>.This spectrometer is very simple, for it has only two scatter plates and a cooled transmission filter to eliminate radiation of which energy is higher than the band gap energy of InSb.The spectral resolutions of this spectrometer are 0.3 cm<SUP>-1</SUP> at 40 cm<SUP>-1</SUP> and 0.4 cm<SUP>-1</SUP> at 90 cm<SUP>-1</SUP>, respectively, which were measured under the constant magnetic field and with sufficiently higher S/N ratio and faster wavelength scanning time than those of conventional far-infrared grating spectrometers.Furthermore, the absorption spectra of water vapor are obtained by taking a tune between the maximum photoresponse of the InSb detector and the scanning of grating.
- 社団法人 日本分光学会の論文
著者
関連論文
- 大学教育としての物理学(討論会)
- 19M-4 Zn,Cd,PbのS,Se,Te化合物の遠赤外域の光学的性質
- 9A23 遠赤外域に於ける二、三の半導体結晶の光斈的性質の測定
- 17L-11 遠赤外分光器の試作
- 17D19. Echelette gratingのスペクトル強度分布について
- レーザ
- 19G-11 InSbの20〜200μの遠赤外域に於ける光学的性質と自由電子の濃度・易動度
- 分光機器の発達
- 最近の遠赤外分光
- 術単光束普通赤外域分光装置
- 波長選択性をもつInSb検知器を用いた遠赤外回折格子分光装置の試作
- 遠赤外干渉分光装置の雑音とデータ処理
- 分析線対選定についての―問題
- 遠赤外干渉分光装置の試作
- 最近の遠赤外分光装置
- Density measurement of plasma by CO2 Laser Interferometer