ピペット取り扱いの安全対策 : 能率の良いピペッター類の試作
スポンサーリンク
概要
- 論文の詳細を見る
The simple pipet filling devices, with which (1〜100) ml pipets can be handled safely, accurately, cleanly, and more quickly, were devised. "Push button pinch clamp type" (Fig, 1) : The device is held in the usual way, sealing the top hole, A, with the right index finger. The liquid is automatically sucked up by lightly pushing the button, F, with the thumb to open the pinch clamp, C. Adjusting to the mark and free drainage are performed rapidly by opening the hole, A, in the usual way. "Nozzle valve type" (Fig. 2) The rubber body, D, is pressed lightly to open the nozzle valve, C and the rubber bulb, E, is kept compressed. The device is held in the usual way, sealing the top hole, A. The liquid is automatically sucked up by pressing the body, D. Adjusting to the mark and free drainage are performed in the usual way. "Finger stall type" (Fig. 3) : It is quite useful particularly for bacteriological or clinical purposes where each sample should be handled by a different pipet. The device is put on the index finger. The finger is brought down on to the top of the pipet so that the holes in the pad and the pipet coincide. As the pad is pressed down lightly, the nozzle valve, C, is opened and the liquid is automatically sucked up. When the liquid has been sucked up, the finger is moved so that the top of the pipet is sealed with the finger tip. Adjusting to the mark and free drainage are performed in the usual way. Pipetting work by using these devices was performed apparently more effectively than that by sucking by the mouth and they may be recommended for chemists, who daily use pipets, in order to make their experiments more comfortable.
- 社団法人日本分析化学会の論文
- 1980-10-05
著者
関連論文
- 自動再結晶法によるプレドニゾロンの高純度高収率精製と薄層クロマトグラフィー及び示差走査熱量測定によるその純度推定
- 高純度精製のための自動再結晶装置の試作
- 迅速自動固液抽出器の試作
- かくはん式自動固液抽出兼再結晶器
- ピペット取り扱いの安全対策 : 能率の良いピペッター類の試作