Performance of Convenient Film Scanner System for Automatic Counting of Track Etched Pits on PADC Detectors
スポンサーリンク
概要
- 論文の詳細を見る
A commercially available film scanner and image processing software "ImageJ" were satisfactorily used for automatic counting of track etched pits on a PADC detector irradiated with α-rays. The performance of the α-ray counting system with the film scanner was examined by the use of a goniometric α-ray irradiation equipment. Under the etching condition of the soakage in 6.25M NaOH solution at 70°C for 20 hours, the α-ray detection efficiency of the film scanner system was about 0.6 time as small as that of the general system with an optical microscope. It was confirmed that PADC detectors together with the film scanner system could be conveniently used for radon monitoring in atmosphere.
- 日本放射線安全管理学会の論文
著者
-
Murata Isao
Division Of Electrical Electronic And And Information Engineering Graduate School Of Engineering Osa
-
Kato Yushi
Division of Electric, Electronic and Information Engineering, Osaka University, Suita, Osaka 565-0871, Japan
-
Yamamoto Takayoshi
Oarai Research Center, Chiyoda Technol Corporation, 3681 Narita-cho, Oarai, Ibaraki 311-1313, Japan
-
Maki Daisuke
Division of Electrical, Electronic and Information Engineering, Graduate School of Engineering, Osaka University, 2-1 Yamadaoka, Suita, Osaka 565-0871, Japan
-
Sato Fuminobu
Division of Electric, Electronic and Information Engineering, Osaka University, Suita, Osaka 565-0871, Japan
-
Iida Toshiyuki
Division of Electric, Electronic and Information Engineering, Osaka University, Suita, Osaka 565-0871, Japan
関連論文
- Measurement and Analysis of Neutron-Induced Alpha Particle Emission Double-Differential Cross Section of Carbon at 14.2MeV
- Development of Cell Chip Based on Track Detector for Examination of Biological Damage by Alpha Particles
- Advanced Burnup Calculation Code System in a Subcritical State with Continuous-Energy Monte Carlo Code for Fusion-Fission Hybrid Reactor
- New Radiophotoluminescence Glass Dosimeter with Specialized Radiation-Sensitive Surface Layer
- Development of a Compact and User-friendly Ion Irradiation System Controlled Remotely through the Internet
- 1P453 X-ray microbeam system for irradiation of living cells(20. Radiation biology,Poster Session,Abstract,Meeting Program of EABS &BSJ 2006)
- Measurement of Charged-Particle Emission Double-Differential Cross Section of Fluorine for 14.2MeV Neutrons
- Digital Image Analysis of Etch Pit Formation in CR-39 Track Detector
- Response of a Radiophotoluminescence Glass Dosimeter to Low Energy X-rays Necessary for Mammography
- Present and Future Views of Division of Radiation Science
- Radiophotoluminescent Observation of X-ray Microbeam Track in Silver-Activated Phosphate Glass
- Performance of Convenient Film Scanner System for Automatic Counting of Track Etched Pits on PADC Detectors