Response of a Radiophotoluminescence Glass Dosimeter to Low Energy X-rays Necessary for Mammography
スポンサーリンク
概要
- 論文の詳細を見る
New-type radiophotoluminescence (RPL) glass dosimeters, which had thin sensitive layers near their surface, were made for the examination of their response to low energy (15-25 keV) X-rays necessary for mammography. We constructed a laser scanning microscope system for the effective measurement of RPL photons from an X-ray irradiated glass dosimeter. The X-ray sensitive region of the new-type glass dosimeter was near its surface, and moreover a thin X-ray absorption glass filter, whose photon attenuation coefficient was almost the same as the glass dosimeter, was put on the new-type glass dosimeter. These new-type glass dosimeters were irradiated with low energy X-rays and measured results of absorbed dose around their surface layers agreed on the whole with results from photon transport calculations. It was confirmed that RPL glass dosimeters could make a reliable service to the low energy X-ray dosimetry for mammography.
著者
-
Kato Yushi
Division of Electric, Electronic and Information Engineering, Osaka University, Suita, Osaka 565-0871, Japan
-
Yamamoto Takayoshi
Oarai Research Center, Chiyoda Technol Corporation, 3681 Narita-cho, Oarai, Ibaraki 311-1313, Japan
-
Sato Fuminobu
Division of Electric, Electronic and Information Engineering, Osaka University, Suita, Osaka 565-0871, Japan
-
Iida Toshiyuki
Division of Electric, Electronic and Information Engineering, Osaka University, Suita, Osaka 565-0871, Japan
-
Kobayashi Hatsumi
Division of Electrical, Electronic and Information Engineering, Graduate School of Engineering, Osaka University
-
Maki Daisuke
Division of Electrical, Electronic and Information Engineering, Graduate School of Engineering, Osaka University
-
Kato Yushi
Division of Electrical, Electronic and Information Engineering, Graduate School of Engineering, Osaka University
-
Hisakado Satoru
Division of Electrical, Electronic and Information Engineering, Graduate School of Engineering, Osaka University
関連論文
- Development of Cell Chip Based on Track Detector for Examination of Biological Damage by Alpha Particles
- New Radiophotoluminescence Glass Dosimeter with Specialized Radiation-Sensitive Surface Layer
- Development of a Compact and User-friendly Ion Irradiation System Controlled Remotely through the Internet
- 1P453 X-ray microbeam system for irradiation of living cells(20. Radiation biology,Poster Session,Abstract,Meeting Program of EABS &BSJ 2006)
- Digital Image Analysis of Etch Pit Formation in CR-39 Track Detector
- Response of a Radiophotoluminescence Glass Dosimeter to Low Energy X-rays Necessary for Mammography
- Present and Future Views of Division of Radiation Science
- Radiophotoluminescent Observation of X-ray Microbeam Track in Silver-Activated Phosphate Glass
- Performance of Convenient Film Scanner System for Automatic Counting of Track Etched Pits on PADC Detectors