副標準電離真空計VS-1型の高真空側における使用限界
スポンサーリンク
概要
- 論文の詳細を見る
The purpose of this paper is to determine the low-pressure limit due to the soft X-ray effect for the sub-standard ionization gauge of type VS-1 which is specified as the standard in "Calibration Method of Vacuum Gauges" of the Japanese Industrial Standards.<BR>Measurements have been made in a bakable ultra-high vacuum glass system by using such two methods as follows;<BR>(1) Observation of the ion collector current as a function of the grid voltage, <BR>(2) Comparison of the ion collector current in the sub-standard ionization gauge with that in the Bayard-Alpert type gauge.<BR>The results indicate that the photoelectric current leaving the ion collector is about 2.8×10<SUP>-10</SUP> A at the normal operating condition, corresponding to a pressure of 7×10<SUP>-9</SUP> mmHg for nitrogen. It is, therefore, considered that there is an error of 1 % in the sensitivity of this gauge at a pressure of 7×10<SUP>-7</SUP> mmHg. Factors affecting the measurements of the photoelectric current are discussed in detail.