Portable and High-Sensitive Apparatus for Measurement of Environmental Radon Using CR-39 Track Detector
スポンサーリンク
概要
- 論文の詳細を見る
Highly sensitive, electrostatic collection chambers have been developed for low-level radon measurements using CR-39 plastic track detectors. Chambers with outer electrodes shaped as a cone and a hemisphere with lines of electric force that were focused on the center of the collector electrode gave alpha-track distributions similar to a 2-dimensional Gaussian with a FWHM of less than 3 mm. The sensitivities, defined as the track density for a radon exposure of 1 Bq h/m^3, were more than 4 tracks/cm^2 using a chamber with a volume of about 300 cm^3. The dependencies of the sensitivity and the track distribution on the collector voltage and humidity were investigated.
- 社団法人応用物理学会の論文
- 1987-04-20
著者
-
MIYAKE Hiroshi
Department of Orthopaedic Surgery, Kagawa Medical Univeristy, Kinashi Ohbayashi Hospital
-
Oda Keiji
Department Of Nuclear Engineering Kobe University Of Mercantile Marine
-
MICHIJIMA Masami
Department of Nuclear Engineering, Kobe University of Mercantile Marine
-
Miyake Hiroshi
Department Of Biosphere Resources Science Graduate School Of Bioagricultural Sciences Nagoya Univers
-
Miyake Hiroshi
Department Of Nuclear Engineering Kobe University Of Mercantile Marine:(present Address) Crocker Nuc
-
Oda Keiji
Department Of Nuclear Engineering Kobe Univ. Of Mercantile Marine
-
Michijima Masami
Department Of Nuclear Engineering Kobe University Of Mercantile Marine
関連論文
- Changes of total body bone mineral density after parathyroidectomy in hemodialyzed patients
- PIXE Analysis of Water Samples
- Characterization of Dextran Sulfate-Induced Guinea Pig Tracheal Plasma Extravasation
- Allelic variations of the D2 dopamine receptor gene in children with idiopathic short stature
- The Arabidopsis FLAKY POLLEN1 Gene Encodes a 3-Hydroxy-3-Methylglutaryl-Coenzyme A Synthase Required for Development of Tapetum-Specific Organelles and Fertility of Pollen Grains
- Portable and High-Sensitive Apparatus for Measurement of Environmental Radon Using CR-39 Track Detector
- Open Abdomen Management After Massive Bowel Resection for Superior Mesenteric Arterial Occlusion
- BEHAVIOR OF ^Po AND ^Po IN THE VICINITY OF LOW FREQUENCY ELECTROMAGNETIC FIELDS
- Clivus Metastasis From Gastric Signet Ring Cell Carcinoma After a 10-Year Disease-Free Interval : Case Report