水銀電極-電解質溶液界面の微分容量測定によるこん跡界面活性剤の定量
スポンサーリンク
概要
- 論文の詳細を見る
Polarographic determination of surfactants in waters usually depends on the suppression of the maximum current of oxygen reductioh. Although it is possible to determine by this method the surfactants in river waters which contain comparatively many surfactants, it is impossible to determine the trace surfactants in piped water. The purpose of the present paper is to establish a method to determine trace surfactants and to apply the method to piped water and underground water. Differential capacity was measured by a.c, polarographs. Differential capacity at a hanging mercury drop electrode decreases with time owing to the adsorption of surfactants. Thus a differential capacitytime curve was used to determine trace surfactants in waters. The detection limits of this method were 5 nM (converted to the concentration of sodium dodecyl sulfate) and I ppb (converted to the concentration of Triton X-100). The concentration of the surfactants in the underground water was 6.5 nM or 1.5 ppb. The concentration of the surfactants in the piped water was 17 nM or 3.2 ppb.
- 社団法人日本分析化学会の論文
- 1989-06-05
著者
関連論文
- 吸着ストリッピングボルタンメトリーによる河川水中のニッケルの定量
- ストリッピングボルタンメトリーによる河川水中の金属の定量
- Izaak Maurits Kolthoff先生の研究と教育
- 2,2′-ビピリジンおよび1,10-フェナントロリン錯体の滴下水銀電極への吸着
- 硫酸ドデシルナトリウムとドデシルベンゼンスルホン酸ナトリウムの水銀電極への吸着-脱着挙動
- パ-ソナルコンピュ-タを使用する中学校理科における化学分野の学習方法
- アセチルアセトンの滴下水銀電極への吸着
- フローインジェクション法を用いる微分容量-時間曲線測定によるビタミンの水銀電極への吸着-脱着挙動
- 金属イオンの吸着ストリッピングボルタンメトリー
- コバルト(2)-2,2′-ビピリジン錯体の水銀電極への吸着とその分析化学的応用