ディスポザブル医材の放射線滅菌線量の検討(II) : ガーゼ付絆創膏のBioburdenの調査および滅菌線量の算出
スポンサーリンク
概要
- 論文の詳細を見る
The bioburden of the commercial adhesive bandage with gauze pad was investigated in order to estimate the sterility dose setting of this item.The distribution of the contaminated microbes on a strip were determined with 1,258 strips of bandage sampled from 12 lots of the products. It was found that the maximum and minimum of the contaminants per strip was 11 and 0,respectively, and the average per strip was 2.65. Besides, no significant differences in the distibution of the numvers of contaminants per strip among the lots were observed.From strips of 5 lots, 235 isolates were obtained and subjected to the taxonomic tests. Most of them were bacilli, of which Bacillus pumilus was predominant. The D value of each strain of 147 isolates sampled from 3 lots, was determined after the irradiation of the prepared test pieces and the D value distribution with microbial population was calculated. The highest D value, 0.19 Mard was obtained with some strains of B. pumilus and B. megaterium. In the comparison with the D values of indicator microbes, B. pumilus ATCC 27142 loaded on the parts of the bandage or on the test pieces, the maximum D value MAXD 10 was assumed to be 0.22 Mrad. Sterility tests of the intact bandages were also carried out after irradiation at incremental doses.From the results obtained, sterilization dose SD was estimated at 1.76 Mrad according to the formula SD=D×log (No/N), on the other hand, according to four methods described in AAMI recommended practice, SD was estimated at 1.52 Mrad, 1.52 Mrad, 1.28 Mrad and 1.33 Mrad, respectively.In the estimations above, it is supposed that 10^<-6> of sterility assurance level of this bandage is attained at the dose less than 2.0 Mrad.
- 日本医療機器学会の論文
- 1984-07-01
著者
関連論文
- 8. ホローファイバー型ダイアライザーの放射線滅菌 : 指標菌のD値の算出
- 2. 殺菌剤クロルヘキシジンによる汚染廃棄医材の全処理方法の検討(その2) : 汚れによるクロルヘキシジンの不活化
- ディスポーザブル医材の放射線滅菌線量の検討(I) : ディスポーザブル注射筒・注射針のBioburdenの調査および滅菌線量の算出
- 20. ディスポーザブル注射針の放射線滅菌条件の検討(第57回日本医科器械学会大会予稿集)
- 18. 放射線滅菌用指標菌の精製方法とD値(第57回日本医科器械学会大会予稿集)
- 6. 自然排気式高圧蒸気滅菌器使用時の工程確立に関する検討
- 75. ガーゼ・脱脂綿の放射線滅菌線量の算出(第58回日本医科器械学会大会)
- 74. ガーゼ・脱脂綿の簡易な汚染菌数測定方法(第58回日本医科器械学会大会)
- 73. ガーゼ付絆創膏のBioburdenの調査と滅菌線量の算出(第58回日本医科器械学会大会)
- 75.ガーゼ, 脱脂綿の放射線滅菌線量の算出
- 74.ガーゼ, 脱脂綿の簡易な汚染菌数測定方法
- 73.ガーゼ付絆創膏のbioburdenの調査と滅菌線量の算出
- 殺菌剤クロルヘキシジンによる汚染廃棄医材の処理方法について
- ディスポザブル医材の放射線滅菌線量の検討(II) : ガーゼ付絆創膏のBioburdenの調査および滅菌線量の算出
- 101.衛生材料の放射線滅菌法 : 照射した日本薬局方脱脂綿、日本薬局方ガーゼの物理化学的変化
- 101.衛生材料の放射線滅菌法 : 照射した局方ガーゼ, 局方脱脂綿の物理化学的変化
- 42.衛生材料の汚染菌数の測定法について
- 39.衛材の滅菌 III : 月経処理用タンポンの滅菌
- 70. 放射線滅菌法における指標菌の検討
- 医療用具の放射線滅菌における滅菌線量の算出法 (I) : シグモイド型生残曲線の標的モデルに対する適合性
- 79. 修理・保守管理に出される医科器械類の消毒法の検討
- 79. 修理・保守管理に出される医科器械類の消毒法の検討(第57回日本医科器械学会大会予稿集)
- 9. ディスポーサブル金属医療用具の放射線滅菌
- 1. 殺菌剤クロルヘキシジンによる汚染廃棄医材の全処理方法の検討(その1) : 医材に対するクロルヘキシジンの吸着
- 滅菌医療用具等のバリデーションに関する調査
- 防菌防黴学会若手の会主催《1995年夏のセミナー》報告
- 各種衛材の放射線滅菌の検討(滅菌法研究会記録)