気体粘性計測用振動子のオゾンガス暴露による周波数-圧力特性の変化
スポンサーリンク
概要
- 論文の詳細を見る
The effect of exposure to ozone on the frequency shift (ΔF) and pressure (P) of a quartz oscillator was observed to check its resistance against the gas. After every ozone exposure, the ΔF-P characteristic curves shifted toward lower frequencies, indicating that the oscillator lost its repeatability. The oscillator electrode consisted of an Au-Cr film deposited on a quartz substrate; an evaluation using a scanning electron microscope (SEM) showed that each film layer had a thickness of 65 nm. The surface composition of the electrode was analyzed by Auger electron spectroscopy (AES). The AES intensities of Cr and Au showed that Cr oxide was present on top of the Au film layer; we speculate that ozone exposure may proceed the Cr oxidation with higher atomic valence. AES depth analysis showed that a small amount of Cr was detected in the Au film layer; we believe that Cr oxidation could progress by either Cr diffusion through the Au layer or by O diffusion from the surface.
- 一般社団法人 日本真空学会の論文
- 2013-05-10
著者
-
小林 太吉
ブイピイアイ(株)
-
北條 久男
バキュームプロダクツ株式会社
-
黒河 明
独立行政法人産業技術総合研究所
-
伊藤 美香
独立行政法人産業技術総合研究所
-
寺内 信哉
独立行政法人産業技術総合研究所
関連論文
- 2種類の圧力計による大気圧O_2/O_3混合ガスのO_3濃度計測
- 水晶摩擦真空計を用いた2成分混合気体の濃度計測
- 水晶摩擦真空計の高精度・高機能化データ処理の研究
- 気体中の水晶振動子のインピーダンスと周波数の圧力依存性の比較
- シリコンのオゾン酸化
- 気体中の水晶振動子のインピーダンスと周波数の圧力依存性の比較
- 気体粘性計測用振動子のオゾンガス暴露による周波数-圧力特性の変化
- 気体粘性計測用振動子のオゾンガス暴露による周波数-圧力特性の変化
- 水晶振動子型水素漏洩検知器における温度の影響とその補正方法