らせん状のイオン軌道を利用した飛行時間質量分析計の開発
スポンサーリンク
概要
- 論文の詳細を見る
A novel time-of-flight mass spectrometer that utilizes a spiral ion trajectory was developed. In this mass spectrometer, the ions sequentially pass through four toroidal electrostatic sectors and revolve along a figure-eight-shaped orbit on a particular projection plane. Each toroidal electrostatic sector has eight stories, and during multiple revolutions, the ion trajectory shifted perpendicular to the projection plane in every cycle, thereby generating a spiral trajectory. Two prototypes, orthogonal acceleration time-of-flight mass spectrometer (OA-TOFMS) and matrix-assisted laser desorption/ionization time-of-flight mass spectrometer (MALDI-TOFMS) were made and tested. The total flight path lengths were 20 and 17 m for OA-TOFMS and MALDI-TOFMS, respectively. By adopting an ion optical system that has a flight path length more than five times longer than that of the commonly used reflectron ion optical system, the m/z-dependent mass resolving power was reduced while improving the mass accuracy of the mass measurements.
- 2009-10-01
著者
関連論文
- 遅延引き出し法の基礎
- らせん状のイオン軌道を利用した飛行時間質量分析計の開発
- 扇形磁場を用いた飛行時間型質量分析計の開発とクラスター分裂の観測
- 19aWD-4 酸化マグネシウム(MgO)_n^+, (MgO)_nMg^+クラスターイオンの強度の相関と分裂パターン
- 19aWD-3 銀・アルミニウムのクラスターの質量スペクトル
- 19aWD-2 スパッター法により作成した水銀・銀クラスターイオンの分裂パターン
- 27aZ-3 スパッター法により生成したテルルクラスターの分裂
- 27aZ-1 二次イオン質量分析法による水銀-銀複合クラスターの研究
- SIMSによる小さいサイズの2価テルルクラスターの観測
- らせん状のイオン軌道を利用した飛行時間質量分析計の開発
- 高性能マトリックス支援レーザー脱離イオン化飛行時間質量分析計を利用したペプチドマスフィンガープリント法によるタンパク質解析能力の検討