Immunochemiluminometric assay of serum TSH by use of chemilumi analyzer.
スポンサーリンク
概要
- 論文の詳細を見る
An immunochemiluminometric assay has been developed for the measurement of thyroid stimulating hormone (TSH) by use of Chemilumi Analyzer (Kyowa Medex). A study was performed to evaluate the feasibility of this system.Linearity was confirmed up to 205μU/ml. Within-run and between-day precision of the method as indicated by C. V. ranged from 2.5 to 6.6% and 2.3 to 11.6%, respectively. Recoveries of TSH from serum were from 96.9 to 107.2%. For TSH, no interference by bilirubin, hemoglobin, turbidity and ascorbic acid was observed. Excellent correlation between TSH measurements on immunoradiometric assay was observed.In the normal subjects the concentration of serum TSH ranged from 0.45 to 2.20μU/ml.