Atomic absorption spectrometry using a flameless graphite cuvette method to measure serum copper was evaluated for its sensitivity, accuracy, and precision. The following results were obtained;(1) linearity: 0-600 μg/100ml, (2) recovery: 99% (98-103%), and (3) reproducibility, within-run C.V.: 1.38%(99.1μg 100ml), 1.85% (46.8μg/100ml), and day to day C.V.: 3.91% (97.8μg/100ml). Other cations, anions or various blood substances did not interfere the measurements. A blood sample of 0.1ml was enough for this analysis and this amount is smaller compared to the previous methods.The present method yields values for serum copper comparable to those obtained by colorimetric procedures.