Required adjustments for obtaining digital cytology images using telecytology
スポンサーリンク
概要
- 論文の詳細を見る
The quality of digital cytology images obtained using a telepatholgy system (OLMICOS, OLYMPUS OPTICS) was occasionally found to be unsatisfactory. The cause of the poor-quality images was not related to system performance, but to inadequate system adjustments for the imaging of cytology specimens. These inadequacies are listed below:<BR>1) Automatic exposure control selects a low exposure lebel.<BR>2) Inadequate gamma-value curve setting.<BR>3) Insufficient magnification of the relay lens between the microscope and CCD camera