Determination of Mercury by Flow Injection-Atomic Fluorescence Spectrometry Using Chromium(II) Reduction-Cold Vapor Generation.
スポンサーリンク
概要
- 論文の詳細を見る
Chromium(II) ions were produced through the reduction of chromium(III) ions by using a flow-type electrolytic cell, and applied to the cold vapor generation of mercury in the presence of iodide. The atomic fluorescence intensity of the vaporized mercury was determined at 253.7nm on an atomic fluorescence spectrometer coupled with a flow injection device. The chromium(II) reduction system could completely reduce the mercury-iodide complex, HgI42-, to elemental mercury, thus enabling an atomic fluorescence determination. The addition of iodide ions (>10mg l-1 as the final concentration) stabilized mercury ions as HgI42- and showed no memory effect due to the adsorption of mercury on the inner walls of the system. Therefore, ample iodide ions were first added to sample solutions; the formed mercury-iodide complex was then reduced to elemental mercury using an on-line chromium(II) reduction system. The present method can be used to determine mercury at the μg l-1 level in aqueous samples without any interference from coexisting iodide ions at a level as great as 100mg l-1.
- 社団法人 日本分析化学会の論文
社団法人 日本分析化学会 | 論文
- ガスクロマトグラフィーによる微量脂肪酸の定量法
- ゼフィラミンによる微量スズ(IV)の吸光光度定量法 : 分析化学におけるオニウム化合物の適用に関する研究(第29報)
- メチレンブルー比色法と赤外吸収スペクトル法の併用による水中アルキルベンゼンスルホン酸塩の定量
- プルトニウムと共存するウランの電位差滴定
- 微量有機物試料の迅速完全酸化法