A Simple Preparation Method of Thin-Layer Standard Samples with Activated Carbon for the Multi-Element Determination of Airborne Particulate Matter by X-Ray Spectrometry.
スポンサーリンク
概要
- 論文の詳細を見る
Thin-layer standard samples for the X-ray fluorescence (XRF) analysis of airborne particulate matter have been conveniently prepared by impregnating activated-carbon powder with standard solutions of the elements of interest and by drying it. Different amounts of standard solutions of the elements of 1.00mg ml-1 each and a 0.01% surfactant (Tween 85) solution of 1.5ml are added to a fine powder of activated carbon of 5-10mg on a hydrophilized poly(tetrafluoroethylene) membrane filter. The suspension of carbon is dispersed by blowing air from a pump for 2-3min without bubbling and dried by IR irradiation for 10min without filtration. This filter with carbon is impregnated with a 0.02% polycarbonate dichloromethane solution of 0.3ml, followed by air-drying. Inductively coupled plasma atomic emission spectrometry of digestates from the prepared samples gave good recoveries of all the measured elements. The XRF intensities of the samples prepared repeatedly had reasonable reproducibilities. Calibration graphs with good linear relationships were constructed for 21 elements in the range of 0-50μg. The standard samples prepared were successfully used for the multi-element XRF determination of airborne particulate matter, in good agreement with the results of instrumental neutron activation analysis.