Heat-Exchange Calorimetry of the Flow Type.
スポンサーリンク
概要
- 論文の詳細を見る
Heat-exchange calorimetry has been extended to the flow type. Two glass vessels for the sample and a reference are fixed separately in a precision water bath. In each vessel, a thermistor, a heater element, an inlet of the solution and an outlet for any overflow are installed. Heat released in the sample vessel is freely exchanged with the ambient water. The temperature in each vessel is observed by a thermistor associated with a Wheatstone bridge. The amplified output is introduced into an analog computation circuit in order to calculate the total heat effect. A constant related to the heat transfer of vessels was theoretically found to be proportional to the flow rate; it was subsequently confirmed experimentally by changing the flow rate from 30g/min to zero, corresponding to the batch type. Since the constancy of the peristalsis pumps is not satisfactory, the flow rate at every setting is observed by means of time and mass measurements. The necessary evaluations for precision and reproducibility are executed in water by electric heating; successful results were obtained for practical use.
- 社団法人 日本分析化学会の論文
社団法人 日本分析化学会 | 論文
- ガスクロマトグラフィーによる微量脂肪酸の定量法
- ゼフィラミンによる微量スズ(IV)の吸光光度定量法 : 分析化学におけるオニウム化合物の適用に関する研究(第29報)
- メチレンブルー比色法と赤外吸収スペクトル法の併用による水中アルキルベンゼンスルホン酸塩の定量
- プルトニウムと共存するウランの電位差滴定