Measurement of the Taiwan Ambient Trace Gas Concentration by Kilometer-Pathlength Fourier-Transform Infrared Spectroscopy.
スポンサーリンク
概要
- 論文の詳細を見る
A portable open long-path Fourier-transform infrared system (POPFT-IR) operable at pathlengths of up to 961m has been constructed for the detection and measurement of trace gases in ambient air and for photochemical smog formation studies. The long optical path is achieved by using a three-mirror multiple reflection cell with a 15.5m base path. The combination of a retro-reflection mirror with one of three mirrors eliminates any air-turbulence interference. The design, construction, advantage, and use of this novel folded open-path optical system and calibration spectrum for quantitative analysis are described in detail. Preliminary ambient air measurements were made at a variety of locations in Taiwan from urban, suburban and rural to remote areas. The existing trace gases in the background atmosphere were measured as nitrous oxide (N2O), methane (CH4), carbon monoxide (CO) and carbon dioxide (CO2). Major pollutants such as ethylene (C2H4), propylene (C3H6), ammonia (NH3), ozone (O3) and peroxyacetyl nitrate (PAN) were measured in the polluted urban atomosphere.
- 社団法人 日本分析化学会の論文
社団法人 日本分析化学会 | 論文
- ガスクロマトグラフィーによる微量脂肪酸の定量法
- ゼフィラミンによる微量スズ(IV)の吸光光度定量法 : 分析化学におけるオニウム化合物の適用に関する研究(第29報)
- メチレンブルー比色法と赤外吸収スペクトル法の併用による水中アルキルベンゼンスルホン酸塩の定量
- プルトニウムと共存するウランの電位差滴定
- 微量有機物試料の迅速完全酸化法