多孔質膜金電極における陽極溶出ボルタンメトリーによる高濃度アルカリ水溶液中の微量銅イオンの定量
スポンサーリンク
概要
- 論文の詳細を見る
A new anodic stripping voltammetry (ASV) using a microporous membrane electrode has been developed for the on-line monitoring of trace CuII in concentrated alkaline solutions. Microporous membrane electrodes were prepared by plating gold onto membrane filters made of polypropylene (PP). By SEM observations, it was confirmed that the membrane filter was modified by gold fine particles. The microporous membrane electrodes were set in a penetrated-type electrochemical flow-cell, and CuII was deposited while the sample solution passed through the electrode. The concentration of CuII was estimated with the anodic stripping peak currents. The calibration curve showed linearity within a range of 0 – 25 μg L−1 CuII in a 48% KOH solution. The proposed method will become a useful technique for an on-line monitoring system for trace CuII, which has been in urgent demand by the semiconductor manufacturing.
著者
-
水口 仁志
山形大学大学院理工学研究科
-
飯山 真充
野村マイクロ・サイエンス株式会社
-
久松 亨
山形大学大学院理工学研究科物質化学工学専攻
-
渡邉 修
山形大学大学院理工学研究科バイオ化学工学専攻
-
水口 仁志
山形大学大学院 理工学研究科
-
久松 亨
山形大学大学院理工学研究科バイオ化学工学専攻
関連論文
- テトラキス(4-カルボキシフェニル)ポルフィンを用いる逆相薄層クロマトグラフィーによる亜鉛イオンの目視定量
- 液滴試料の逆相薄層クロマトグラフィーによるppbレベルのアルミニウムイオンの選択的スポットテスト
- 薬液用金属除去モジュール(メトレート®)の開発と実用化
- PTFE製メンブランフィルターへのイオン対固相抽出によるクロム (VI) の高感度目視定量
- 科学への純粋な興味
- フロー電解セルを用いるストリッピングボルタンメトリーによる高濃度アルカリ水溶液中の銅(II)とニッケル(II)の同時微量定量
- 高性能化する二重電極ジェネレータ・コレクタモード
- 多検体試料中の総ポリフェノール量の迅速定量のためのスキャナーを用いるフォーリンチオカルト比色分析
- β-ジケトンを付加したケイ酸塩を用いた二価金属イオン間の分離能に関する研究
- 多孔質膜金電極における陽極溶出ボルタンメトリーによる高濃度アルカリ水溶液中の微量銅イオンの定量