UF6用四重極子型質量分析計の開発
スポンサーリンク
概要
- 論文の詳細を見る
The quadrupole type mass spectrometer system for the precise determination of <SUP>235</SUP>U isotope ratio of UF<SUB>6</SUB> has been developed. Fundamental performances, such as inlet and evacuation characteristics, resolution, relative sensitivity, pattern coefficients, isotopeion current detection characteristics were measured and examined carefully on this quadrupole type mass spectrometer for UF<SUB>6</SUB>. The operational stability of this quadrupole type maas spectrometer was good. <SUP>235</SUP>U and <SUP>238</SUP>U isotope peak was resolved perfectly. Secondary electron multiplier tube was effective for rapid measuring low ion current with high efficiency, but it had about 3% systematic error on determination of <SUP>235</SUP>U isotope ratio. It was able to correct this error easily by calibration with standard sample. Results of the performance tests were good. <SUP>235</SUP>U isotope ratio measurements were carried out by use of many UF<SUB>6</SUB> standard samples. The precision and the accuracy of the measurements were less than ±0.3% under appropriate operating conditions attained. The time required for measuring one sample was less than five minutes.
- 日本質量分析学会の論文