電子管制御式窒素自動燃焼装置
スポンサーリンク
概要
- 論文の詳細を見る
A new automatic apparatus for the micro-Dumas method was described.The moving furnace was moved at the rate of 13mm/min.by a synchronized motor and it was stopped whenever the pressure in t e combustion tube was increased.Thus, the furnace did not move until it was heated to the maximum temperature, and it was also stopped when the sample was evaporated or oxidized.Mercury was used in the manometer as shown in Fig.3.Another relay was connected to the electric bubble counter as shown in Fig.3.B, and the volume of the gas passed through was counted. Mercury and a small quantity of liquid paraffin was placed in the bubble counter. A gasometer of 200cc (Fig. 1 Az<SUB>2</SUB>) was equipped to check the counter and to substitute the air with pure CO<SUB>2</SUB> quantitatively.In this apparatus, fine stop cock (Fig.1 k<SUB>2</SUB>) was always set to give the gaseous velocity of 4-5cc/min.The stop cock k<SUB>3</SUB> was closed during the stage of buring of the sample and it was opened during the sweeping stage.
- 社団法人 日本分析化学会の論文
著者
-
河野 哲郎
東京大学農学部農芸化学科
-
中山 道子
東京大学農学部農芸化学科
-
三井 武子
東京大学農学部農芸化学科
-
佐藤 惠子
東京大学農学部農芸化学科
-
河野 哲郎
東京大学農学部農芸化学教室
-
中山 道子
東京大学農学部農芸化学教室
関連論文
- 酸素の微量分析 : (第6報)長期試験成績
- 電子管制御式硫黄及びハロゲン分析装置
- 自動式Micro DUMASの改良
- 酸素の微量分析 : (第1報)予備実験及び窒素の精製条件
- 電子管制御式窒素自動燃焼装置
- 電子管制御式炭水素自動燃焼装置
- 電子管制御法による微量元素分析の総括 : 長期試験及び綜合分析の結果
- 高周波を利用したフラクションコレクター用リレー : 高周波による化学反応の追跡第4報
- 高周波による化学反応の追跡 : (第2報)滴定曲線の解釈及びGlyci eの滴定曲線
- 高周波による化學反應の追跡 : (第1報) Micro-KJELDAHLの滴定への応用
- 酸素の直接微量分析法
- 炭水素分析値の季節的変動について
- 酸素の微量分析 : (第2報)熱分解の自動制御装置
- タイトル無し
- アスカライトの調製