An accurate flowmeter based on the reduced constant-pressure method was devised for admitting the sample gas into the gauge calibration system. This flowmeter consists of a U-tube manometer, an accurate micrometer for measuring the gas pressure, and a diaphragm gauge for checking the deviation of the gas pressure. The gas pressure was kept constant by raising the mercury level in the U-tube manometer. The flow rate could be obtained from the gas pressure, the cross-sectional area of the U-tube manometer, and the rise-up rate of mercury level. With the U-tube manometer of 34. 0±0.1 mmφ in diameter, the flow rate of 10<SUP>-6</SUP> to 10<SUP>-2</SUP> Torr·<I>l</I>/s could be easily measured with an accuracy better than 1%. An actual example in which this flowmeter was used for the gauge calibration was presented.