Development of an in vivo tissue-engineered, autologous heart valve (the biovalve) : preparation of a prototype model
スポンサーリンク
概要
- 論文の詳細を見る
Objective: This study aimed to develop an autologous heart valve without using traditional in vitro tissue-engineering methods, which necessitate complicated cell management protocols under exceptionally clean laboratory facilities. Methods: An autologous heart valve construct composed of trileaflets was prepared using a specially designed mold. The mold was prepared by covering a silicone rod with a crown-shaped tubular polyurethane scaffold containing 3 horns. The mold was implanted in the dorsal subcutaneous space in Japan White rabbits for 4 weeks. After harvesting, the implanted trileaflet valve-shaped structure with an internal diameter of either 5 or 20 mm was obtained by trimming the membranous tissue formed between the horns located around the silicone rod. The valve substitute was examined both macroscopically and histologically. The tensile strength of the leaflets was measured to rupture. The degree of regurgitation in valve function was evaluated using a flow circuit by calculating the ratio of the regurgitation volume to the forward flow volume. Results: After implantation, the mold was completely covered with connective tissue consisting mostly of collagen and fibroblasts. Harvesting of the mold was straightforward, because there was little adhesion between the formed tissue and the native skin tissue. The trileaflet heart valve construct was obtained after withdrawing the inserted rods and trimming the membranous tissues formed between the horns of the scaffold. It was firmly attached to the scaffold, the interstices and surface of which revealed connective tissues composed of components similar to those of the leaflet tissue. Although the mechanical properties of the leaflet tissue were less efficient than those of the native porcine aortic valve leaflets, satisfactory valvular functions were demonstrated under pulsatile conditions using a flow circuit. No regurgitation was observed under retrograde hydrostatic pressures of up to 60 mm Hg, the physiologic pressure acting on the aortic valves during retrograde aortic flow. Conclusions: The biovalve, an autologous, in vivo tissue-engineered, trileaflet, valve-shaped construct, was developed using our novel in-body tissue architecture technology. The biovalve has the potential to be an ideal prosthetic heart valve, with excellent biocompatibility to the growth of the recipient’s heart.
- Mosby, Inc.の論文
著者
-
Ando Joji
獨協医科大学 医学部生体医工学
-
KANDA Keiichi
Department of Cardiovascular Surgery, Kyoto Prefectural University of Medicine
-
Kanda Keiichi
Department Of Cardiovascular Surgery Kyoto Prefectural University Of Medicine
-
Nemoto Yasushi
Chemical Products Development Department Bridgestone Corporation
-
Nakayama Yasuhide
National Cardiovascular Center Res. Inst. Osaka Jpn
-
Nakayama Yasuhide
Department Of Bioengineering Advanced Biomedical Engineering Center National Cardiovascular Center R
-
Ando Joji
Lab. Of Biomedical Engineering School Of Medicine Dokkyo Medical Univ. Jpn
関連論文
- Histological Evaluation of FK506 Eluting Polymer Covered Stents in Rabbits
- Laser perforated collagen membrane : pore size-dependent bone induction as a new BMP carrier
- Local elasticity imaging of vascular tissues using a tactile mapping system
- Development of an in vivo tissue-engineered, autologous heart valve (the biovalve) : preparation of a prototype model
- Development of biotube vascular grafts incorporating cuffs for easy implantation
- Development of sutureless vascular connecting system for easy implantation of small-caliber artificial grafts
- Differentiation from embryonic stem cells to vascular wall cells under in vitro pulsatile flow loading
- High-performance self-expanding stent graft : development and application to experimental aneurysms
- Local Delivery of Low-Dose ONO-4817, a Broad-Spectrum Matrix Metalloproteinase Inhibitor, Reduces Neointimal Hyperplasia Coupled with Decreased Collagen Deposition
- Local Drug Delivery for Prevention of Post-Angioplasty Restenosis
- Development of in vivo tissue-engineered autologous tissue-covered stents : biocovered stents
- Influence of vascularized transplant bed on fat grafting
- Adequate Design of Customized cDNA Macroarray for Convenient Multiple Gene Expression Analysis(METHOD)
- Development of novel covered stents using salmon collagen
- Photodimerization and Polymerization of PEG Derivatives through Radical Coupling using Photochemistry of Dithiocarbamate
- Surface elasticity imaging of vascular tissues in a liquid environment by a scanning haptic microscope
- Preparation of in-vivo tissue-engineered valved conduit with the sinus of Valsalva (type IV biovalve)
- Yield Stress of Randomly Perforated Sheets and Anisotropic Yield Function
- In vivo evaluation of poly(N-isopropylacrylamide) (PNIPAM)-grafted gelatin as an in situ-formable scaffold
- Material design for an artificial extracellular matrix: cell entrapment in poly (N-isopropylacrylamide) (PNIPAM)-grafted gelatin hydrogel
- Effectiveness of Interactive Class with Computer Assisted Visual Presentation
- In-body optical stimulation formed connective tissue vascular grafts, "biotubes," with many capillaries and elastic fibers
- Development of microporous self-expanding stent grafts for treating cerebral aneurysms : designing micropores to control intimal hyperplasia
- Variations in local elastic modulus along the length of the aorta as observed by use of a scanning haptic microscope (SHM)