1132 近赤外線水分計を用いた軟骨含水量の測定(第2報) : 正常関節軟骨の検量線作成(S11-2 計測と力学-生体への応用(2),S11 計測と力学-生体への応用)
スポンサーリンク
概要
- 論文の詳細を見る
In previous studies, we showed that it was able to determine mechanical properties of cartilages by measuring NIR absorbance after indentation test. In this study, calibration curves for gelatins and human articular cartilages were obtained to estimate their water content quantitatively. NIR absorbance and water content in gelatin and human articular cartilages were measured, and their relations were discussed. Gelatin showed NIR absorbance was proportion to water content. Linear relations between water content per thickness and thickness and NIR absorbance and between thickness and NIR absorbance were found in human articular cartilage. These results enabled us to estimate water content in both gelatin and cartilage by the NIR absorbance. The discrepancy between measured and estimated water content in cartilage was 2.95%.
- 一般社団法人日本機械学会の論文
- 2005-09-18