Driven Equilibrium Pulseを利用したTurbo Spin Echo法(DRIVE)の基礎的検討
スポンサーリンク
概要
- 論文の詳細を見る
When MR images are obtained with the turbo spin echo (TSE) sequence, DRIVE can be used as a sequence in which the driven equilibrium pulse (DE pulse), a reset pulse, is applied at the TSE echo train to accelerate relaxation time and return to the equilibrium of Mz magnetization. In this study, we examined the extent to which TR could be shortened in DRIVE and how the other parameters of the turbo spin echo sequence influence it. Results: 1) DRIVE is effective when the T2 value is long. 2) It is necessary to set TR at 1000ms or more to obtain image contrast with free water and fat in T2-weighted images for which a conventional turbo spin echo sequence using DRIVE is employed in clinical examination. 3) It is not necessary to consider the influence of the TSE factor when using DRIVE.
- 公益社団法人日本放射線技術学会の論文
- 2006-07-20
著者
-
井上 雄介
京都第一赤十字病院・放射線科
-
金澤 裕樹
京都第一赤十字病院・放射線科
-
井上 雄介
京都第一赤十字病院放射線科
-
金澤 裕樹
Department Of Radiology Kyoto First Red Cross Hospital
-
石田 憲一
京都第一赤十字病院第一放射線科
-
石田 憲一
京都第一赤十字病院放射線科
関連論文
- 134 動態曲線の信号強度測定による乳腺腫瘍組織型推定の試み(MR検査 乳腺,一般研究発表,日本放射線技術学会 第34回秋季学術大会)
- 高速スピンエコー法を用いたT1・T2同時計測MRI
- 高速スピンエコー法を用いた造影ダイナミックT1値計測MRIの最適条件の検討
- 337.乳房2倍拡大撮影法(第3報) : カーボン・ファイバー・カセッテについて(診断用装置-8 CFRP他)
- Driven Equilibrium Pulseを利用したTurbo Spin Echo法(DRIVE)の基礎的検討
- 平成20年度後期国際研究集会派遣会員報告書
- 平成20年度後期国際研究集会派遣会員報告書 : RSNA2008参加報告書