Energy Analyser in the Axial Magnetic Field for Hot Electrons Emitted from Plasmas
スポンサーリンク
概要
- 論文の詳細を見る
Two new modifications of the energy selective analyser working in an axial magnetic field and with a perpendicular electric field are proposed. Their characteristics are discussed in view of application for the plasma diagnostics, especially for the energy measurement of the particles escaping from a linear system. The energy resolution of one of them is calculated in detail and one design of the analyser with the multichannel detectors is shown. A brief description of its construction and a preliminary result of the calibration is also given.
- 核融合科学研究所の論文
著者
関連論文
- A Quick Acting Gas Valve Driven through a Resistive Wall
- A Simple Vacuum Crowbar Switch up to 1 MA
- Electron Beam Extraction from INS AG Synchrotron. : Part II. Experiment
- Coaxial Marx Generator for Producing Intense Relativistic Electron Beams
- A High Resolution Echelle Mortochrornator and Its Application to Ion Temperature Measurement of He Plasma
- Electron Beam Extraction from INS AG Synchrotron. : Part I. Theory
- Energy Analyser in the Axial Magnetic Field for Hot Electrons Emitted from Plasmas
- Energy Analyser in the Axial Magnetic Field for Hot Electrons Emitted from Plasmas
- Diffusion, of, Multipole, Magnetic, Fields, through, a, Conductive, Wall, of, a, Plasma, Container