Study of the Magnetic Field Effect on Commercial Thermistors using a Water Triple Point
スポンサーリンク
概要
- 論文の詳細を見る
The effect of a magnetic field on commercial thermistors was precisely studied in detail using the water triple point. A thermistor with a code of 44004 manufactured by YSI Inc. was found to show a small correction. Its orientation effect and sensor dependence was found also to be small. The correction $\Delta T$ (mK) against magnetic field $B$(T) was found to be 0.069($B$/T)2 (mK). The uncertainty of the correction function is estimated to be as small as 2 mK up to 15 T.