Improvement of a Dose Reading System for Radiophotoluminescence Glass Dosimeters
スポンサーリンク
概要
- 論文の詳細を見る
A new dose reading system for radiophotoluminescence(RPL) glass dosimeters was constructed by the use of a highly repetitive Q-switched solid state laser in place of a conventional nitrogen gas laser. It was confirmed that the high repetition rate of laser pulses for the excitation was effective for the improvement of the signal-to-noise ratio in the RPL measurement and eventually led to the improvement of the sensitivity of the dose reading system. In addition, several optical improvements were made for the avoidance of the annealing effect in the RPL measurement. The performance tests with normal RPL glass dosimeters indicated that the new dose reading system had high sensitivity and good long-term stability in radiation dose measurement.
- 公益社団法人 日本アイソトープ協会の論文
公益社団法人 日本アイソトープ協会 | 論文
- Improvement of a Gas Scintillation Proportional Counter
- Questionnaire Survey of SPECT Studies after the Supply Shortage of Tc-99m and Model Analysis Concerning Earnings and Expenditure of Generator Use
- 各種骨病変検出における臨床的有用性と医療経済効果に関する骨シンチグラフィと`18´F-fluoride PETの比較検討
- Combination of Sulfur Isotope Ratio of Non-sea Salt Sulfate and Lead-210 Concentration in Aerosols as an Index of Long-range Transported Aerosols
- 放射線生物学の動向