Microbiological Challenge Test of Contamination Caused by Using the PhaSeal System
スポンサーリンク
概要
- 論文の詳細を見る
In the Japanese Pharmacopoeia, there are no criteria for multiple uses, such as multidose vials, for rubber stoppers (RS), which may cause bacteriological contamination. To resolve this problem we would like to suggest the use of a closed system (CS), while considering the structure. However, multidose vials are not always kept in an aseptic environment. Hence, it is necessary to assess CS contamination and metal needles, so we compared the microbiological contamination level using Bacillus subtilis between the PhaSeal system (PS) and metal needles.On the RS of a vial containing distilled water and the top of PS-P21s membrane, 3.0 × 104 spores were prepared ("Bacillus-vial" and "Bacillus-P21"). An unused P21 was attached to the Bacillus-vial (Sample1). A metal needle was vertically inserted into the Bacillus-vial, and each of the puncture and vial inversion operations was carried out once, 5 times, and 10 times (Sample 2-1, 2-5, 2-10). Furthermore, Bacillus-P21 was attached to the Bacillus-vial, and PS-N35 was attached to Bacillus-P21. Each puncture was carried out once, 5 times, and 10 times (Sample 3-1, 3-5, 3-10). Viable bacterial cells were counted using the pour plate method.Bacterial cells were detected in all samples; however, bacterial cells were not detected from control vials that were not administered prepared spores. The B. sublitis contamination rates of Sample 1/2-1/2-5/2-10/3-1/3-5/3-10 were 0.7/1.4/3.1/1.9/0.1/3.2/5.0 %, respectively.This study suggests that PS is almost equivalent to metal needles at up to 5 punctures in terms of microbiological contamination when PS is used multiple times for multidose preparation under preparation conditions where the vials are susceptible to bacterial contamination.
- 日本医療薬学会の論文
日本医療薬学会 | 論文
- O17-003 視覚障害者に対する手技指導の標準化を目的とした視覚障害擬似体験インスリン注射院内実習(一般演題 口頭発表,糖尿病/後発医薬品,医療薬学の創る未来 科学と臨床の融合)
- 20H-16 Basal-Bolus療法におけるインスリンデテミルの有効性(使用状況調査・意識調査,来るべき時代への道を拓く)
- 30-P2-43 2種類のインスリン試薬を用いた化学発光酵素免疫測定法によってFactitious hypoglicemiaを証明しえた一例(副作用・有害事象,社会の期待に応える医療薬学を)
- 28-04-20 ペン型インスリン注射器の精度に及ぼす注入時抵抗とカートリッジ内気泡の影響
- P-610 ファブリー病に対する「ファブラザイム^[○!R]」投与時の「過敏反応発現時の対処方法」の一部改変により副作用防止に至った1症例(8.有害事象・副作用(基礎と臨床)1,医療薬学の未来へ翔(はばた)く-薬剤師の薬剤業務・教育・研究への能動的関わり-)