容積補償型連続血圧測定のための加圧振動法によるサーボ制御目標値の短時間決定法
スポンサーリンク
概要
- 論文の詳細を見る
To perform the instantaneous radial blood pressure measurement based on the volume-compensation method, a reference value for the servo control (V0) has to be determined in the initial stage of the measurement. The volume-oscillometric method with a photo-plethysmography has previously been used to determine the V0. However, an operational time by this method requires more than 30 s. In order to reduce this time and/or to reset quickly the V0 when it suddenly changes during the measurement, we have developed a new method to determine the V0 more quickly. In this new method, sinusoidal vibrations having a frequency of 20 Hz and an amplitude of 10 mmHg are superimposed on the cuff pressure that is increasing by the rate of 20 mmHg/s. The V0 can be determined by detecting a mean value of the photo-plethysmographically measured arterial volume showing the maximum volume amplitude corresponding to the pressure vibration component. The validity of this method was evaluated using twenty healthy subjects by comparing with the volume-oscillometric method as a reference. The results indicated that the present method could determine the V0 with very high accuracy: The average of the ratio of V0 determined by these two methods was 1.002 with standard deviation of 0.012. This paper also described the effects of the levels of blood pressure, pulse rates and the circumferences of the wrists on the determination of V0, together with the resetting of V0 during the instantaneous blood pressure measurement when the initial V0 was suddenly changed due to an artificial operation.
- 一般社団法人 日本生体医工学会の論文
一般社団法人 日本生体医工学会 | 論文
- Evaluating the Workload Reduction of Automatic Vital Data Transmission
- Mental Fatigue Measurement Based on the Changes in Flicker Perception Threshold using Consumer Mobile Devices
- A Theoretical Study on a Computational Algorithm for Human Posture Estimation Based on Motion Capture of a Small Number of Markers
- Latest microscopic technique. From the principle to application. Development of three-dimensional microscopic system for operation of mini-pig fertilized eggs.
- A Study of the Automation of Cytodiagnosis (The Second Report):Cytophotometry Using Photomicroscanner and a Trial of Data-processing