Low-background shielding for a well-type Ge detector.
スポンサーリンク
概要
- 論文の詳細を見る
<I>A lead-shielding was constructed for a welltype Ge detector to utilize low-level r spectrometry. Background counting rate was investigated under lead thickness of 5-25 cm. It has been confirmed that selection of low activity materials for the construction of the detector has advantage to reduce the contribution of background deu to natural radioactivities.</I> N<SUB>2</SUB> <I>gas was helpful to reduce the contribution of atmospheric radon daughters especially during ofd ventilation period.</I>