<I>Organ volume was experimentally and clinically determined using single photon emission computed tomography (SPECT) . In organ phantom filling with</I> <SUP>99<I>m</I></SUP>TcO<SUB>4</SUB><SUP>-</SUP>, <I>the error was less than 3.3% in the volume over 400 ml but 11% at 200 ml. In 28 cases of patient, liver and spleen volume were measured by SPECT and X -CT. Good correlation was founded between SPECT volume and X-CT volume (r=0.99, n=49), although SPECT volume were 16% higher than X-CT volume (y=1.16 x+11.6, x: X-CT volume, y: SPECT volume) </I>.