タイトル無し
スポンサーリンク
概要
- 論文の詳細を見る
A new type of valve system for surgical treatment of hydrocephalus has been designed and developed. Results obtained from basic experiments by a scale-up model valve are summarized as follows.<BR>1. Regarding the pressure control of the valve, the opening action of the valve occurs at the critical pressure difference, the flow through the valve begins at the pressure more than this critical one. The increase of the pressure difference along with increase of the flow is comparatively small even of large flow.<BR>2. The prevention of counter flow is almost satisfactory with this valve. 3. With this valve, the cerebrospinal fluid pressure is easily controlled and hence, the valve may be applied to all patients with the abnormal cerebrospinal fluid pressure.<BR>4. A significant advantage of this valve is the existence of the popping pressure, compared with other resistance-type valves.
- 一般社団法人 日本生体医工学会の論文
一般社団法人 日本生体医工学会 | 論文
- Evaluating the Workload Reduction of Automatic Vital Data Transmission
- Mental Fatigue Measurement Based on the Changes in Flicker Perception Threshold using Consumer Mobile Devices
- A Theoretical Study on a Computational Algorithm for Human Posture Estimation Based on Motion Capture of a Small Number of Markers
- Latest microscopic technique. From the principle to application. Development of three-dimensional microscopic system for operation of mini-pig fertilized eggs.
- A Study of the Automation of Cytodiagnosis (The Second Report):Cytophotometry Using Photomicroscanner and a Trial of Data-processing