Simultaneous Determination of Inorganic As(III), Sb(III) and Se(IV) by HG/ICP-AES using Citric/Citrate Buffer.
スポンサーリンク
概要
- 論文の詳細を見る
A method has been developed for the simultaneous determination of inorganic As (III), Sb (III) and Se (IV) by hydride generation inductively coupled plasma-atomic emission spectrophotometry (HG/ICP-AES), based on the use of citric citrate buffer.<BR>In citric 'citrate buffer at pH 3.0, it was possible to inhibit the reduction of As (V), Sb (V) and Se (VI) to AsH<SUB>3</SUB>, SbH<SUB>3</SUB> and H<SUB>2</SUB> Se and hence to selectively determine As (III), Sb (III) and Se (IV) . This method allowed detection limits of 0.2 ng As (III) /ml, 0.2 ng Sb (III) /ml and 1.4 ng Se (IV) /ml and quantitative limits of 1 ng As (III) /ml, 1 ng Sb (III) /ml and 4 ng Se (IV) /ml. The calibration curves of As (III) and Sb (III) were liner from 1 to 100 ng/ml, while that of Se (IV) was liner from 4 to 100ng/ml.<BR>The method can be applied to determine As (III), Sb (III) and Se (IV) in river water, ground water and soil solution.