A method for simultaneous analysis of pesticides in the atmosphere by means of GC/MS using activated carbon fiber filter paper was examined. The target compounds were 43 pesticides including aerially applied ones. Air was passed at a flow of 10l·miri<SUP>-1</SUP> through two activated carbon fiber filter papers with a diameter of 47 mm (Toyobo KF Paper P-175) for 7 days. Each sampled filter was extracted ultrasonically with 10 ml of toluene/ethanol (4: 1) for 15 min followed by centrifugation at 3000 rpm for 15 min. The extraction was repeated 3 times. The extracted solution was concentrated to almost dryness, and 1 ml of acetone was added. This acetone solution was analyzed quantitatively with GC/MS. The detection limit for this method was from 0.0001 μg·m<SUP>-3</SUP> to 0.002 μg·m<SUP>-3</SUP>. In the recovery test, recovery rates were 70-109 % for 36 pesticides. Using this method, variation of pesticides in the atmosphere was monitored in a rural area in Niigata Prefecture. Twenty-two pesticides were detected in the atmosphere at 0.0001 -0.041 μg·m<SUP>-3</SUP> in median. This method has been successfully used for the monitoring of pesticides in the atmosphere.