Comparison of Methods for Estimating Limit of Detection in GC/HRMS for Dioxins.
スポンサーリンク
概要
- 論文の詳細を見る
Five methods for estimating the limit of detection are compared with regard to the estimates of LOD and reproducibility of the LOD estimates in GC/HRMS for dioxins. The following methods are taken:<BR>1. LOD based on repetition:<BR>Three times SD of area measurements (area SD, n = 5) ;<BR>Three times SD of height measurements (height SD, n=5) ;<BR>2. LOD based on S/N:<BR>Six times SD of baseline noise (baseline SD) ;<BR>Six fifths of envelope of baseline (baseline range) ;<BR>3. LOD based on probability theory of signal and noise (FUMI theory) .<BR>The LOD estimates obtained from the above methods, except the height SD, are comparable to each other. The reproducibility of the LOD estimates is ca. 30 % RSD for the area SD, height SD and baseline range, ca. 20 % RSD for baseline SD and ca. 15 % RSD for the FUMI theory. The FUMI theory is concluded to be the best choice from the viewpoints of statistical reliability and practicality.