A cassette-tape long-term (up to 24 hours) ECG recorder and high-speed analyzer for the tape has been designed and developed experimentally. By using modern electronics technology, such as integrated circuits, shift register and phase shiftless filter, the developed system has proved to have characteristics which are more or less satisfactory. The high-lights of the system : i) the heart-rate is given by each beat-to-beat interval rather than by an averaging of a certain period, ii) -recording and analysis are carried out by two channels rather than one, iii) phase shiftless filter is utilized, and iv) ST deviation is detected and recorded on time scale. By using the system, 78 recording and analysis sessions were held with clinical cases and with healthy subjects as well, and the usefulness of the system was proven in detecting and counting episodes including extrasystoles, atrioventricular block and ST segments deviation. Problems encountered and possible means for their solution, too, are described.