Evaluation of the Rapid Measurement Kit "SimPlate" of Total Count, Coliform Count and E. coli Count in Food Samples
スポンサーリンク
概要
- 論文の詳細を見る
Total plate counts, coliform counts and <I>E. coli</I> in food samples determined by a rapid measurement kit " SimPlate" were compared with those determined by conventional methods: the pour plate methods using plate count agar, desoxycholate agar and 3M Petrifilm EC plate. The results are shown below.<BR>1. The correlation coefficient between the total counts in 154 food samples determined using "SimPlate" and the standard agar plate method was 0.95, showing a very high correlation.<BR>2. Of 258 food samples, 131 and 112 samples were coliform positive on plates using "Sim-Plate" and desoxycholate agar. The correlation coefficient of the coliform counts on the two media was 0.79, showing a good correlation.<BR>3. Of 65 pork meats, <I>E. coli</I> was detected, using "SimPlate" and 3M Petrifilm EC plate 54 and 52 samples, respectively, and the correlation of <I>E. coli</I> counts obtained by the two methods was 0.86.<BR>4. 96 strains isolated from fluorescent wells of "SimPlate" were <I>E. coli</I> 95 strains and <I>K. cryocrescens</I> 1 strain.<BR>The above findings clarified that "SimPlate" obtained results comparable to those obtained by the conventional total count, coliform counts and <I>E. coli</I> count methods in food samples and thus, the rapid measurement kit "SimPlate" is useful for rapid on-site testing of food samples.