Evaluation of the rigidity of dentures made of injection-molded materials
スポンサーリンク
概要
- 論文の詳細を見る
Dentures made of 2 different types of injection-molded thermoplastic resins (polyamide resin and polyester resin) and a denture made of conventional heat-polymerized resin were used to create an experimental model of a mandibular molar region with a two-tooth gap. In the experimental model, a force of 100 N was applied onto the mesial fossa of the first molars of the dentures, and comparisons were performed by measuring the pressure applied under the denture base and the subsidence rate of the denture. The polyamide resin denture showed the highest subsidence, exerted the highest pressure on the underlying mucosa, and showed significant differences with the other types of dentures. The findings showed that polyamide resins have the lowest degree of elasticity, and that when resins with such low elasticities are used in the denture base, they should preferably be reinforced with metals.
著者
-
Sato Masayuki
Removable Partial Denture Prosthodontics Department Of Masticatory Function Rehabilitation Division
-
Igarashi Yoshimasa
Removable Partial Denture Prosthodontics Department Of Masticatory Function Rehabilitation Division
-
WADACHI Juro
Removable Partial Prosthodontics, Masticatory Function Rehabilitation, Division of Oral Health Sciences, Graduate School, Tokyo Medical and Dental University
-
IGARASHI Yoshimasa
Removable Partial Prosthodontics, Masticatory Function Rehabilitation, Division of Oral Health Sciences, Graduate School, Tokyo Medical and Dental University
関連論文
- Comparisons of Masticatory Movements between Mixing Ability Test and Masticatory Performance Test
- Evaluation of the static frictional coefficients of Co-Cr and gold alloys for cone crown telescope denture retainer applications
- Analysis of the Relationship between the Incisal Overjet in a Maxillary Denture and Phonetic Function Using a Speech Recognition System
- Effect of reactive adhesives on the tensile bond strength of polyvinyl siloxane impression materials to methyl methacrylate tray material
- Reinforcement effects of fiberglass on telescopic dentures using a three-dimensional finite element analysis and fracture test
- A follow-up study on removable partial dentures in undergraduate program:Part I. Participants and denture use by telephone survey
- Evaluation of Clinical Training for Removable Partial Denture at the Tokyo Medical and Dental University
- Effect of mold temperatures on interface between primary and secondary castings of cast-on method for precision metal frameworks
- A Questionnaire Study of Japanese Dentists' Perceptions of Risks for Overeruption of Molars without Antagonists and Attitude on Prosthetic Intervention
- Comparison of electromyographic activity of jaw-closing muscles between mixing ability test and masticatory performance test
- Effects of surface roughness and tapered angle of cone crown telescopic system on retentive force
- Bilinear elastic property of the periodontal ligament for simulation using a finite element mandible model
- Clinical Evaluation of Semi-Adjustable Articulators : Reproducibility of Sagittal Condylar Path Inclination Assessed by a Jaw-Tracking System with Six Degrees of Freedom
- Evaluation of the rigidity of dentures made of injection-molded materials
- Analysis of the Relationship between Palatal Contour and the Phonetic Function in Complete Denture Wearers Using a Speech Recognition System
- Comparison of electromyographic activity of jaw-closing muscles between mixing ability test and masticatory performance test
- Influence of removable partial dentures on the formation of dental plaque on abutment teeth
- Evaluation of the rigidity of dentures made of injection-molded materials
- Influence of the major connector in a maxillary denture on phonetic function