LC-MSを用いるフェニルヒ素化合物の分析条件の検討と茨城県神栖市の地下水試料への適用
スポンサーリンク
概要
- 論文の詳細を見る
An analytical method for the determination of Diphenylarsinic acid (DPAA), Phenylarsonic acid (PAA) and Phenylmethylarsinic acid (PMAA) in groundwater by liquid chromatography mass spectrometry (LC/MS) has been developed. The phenylarsenic compounds were separated by LC with an anion-exchange column, and identified by electrospray ionization mass spectrometry using the m/z 261, 201, and 199 as monitoring ions for DPAA, PAA and PMAA, respectively. Quantitative measurements were performed by the internal-standard method with isotope-labeled phenylarsenic compounds as internal-standard compounds, in which quantification was not influenced by the matrix effect. The instrument quantification limits by this analytical method were 0.77 μg L−1, 0.48 μg L−1 and 0.30 μg L−1 for DPAA, PAA and PMAA, respectively. We found monitored phenylarsenic compounds in groundwater by this method for two years. The occupancy of each phenylarsenic compound to the total arsenic concentration was 85% (DPAA), 11% (PAA) and less than 1% (PMAA), respectively, until the earthquake that occurred on 11th Mar 2011, but that varied after the earthquake. The cause was presumed to be due to the exchange of the pumping wells.
著者
関連論文
- 不正軽油製造に伴う廃棄物中のクマリンの定量
- 不正軽油製造に伴う廃棄物中のクマリンの定量
- 固相抽出-液体クロマトグラフィー/質量分析法による環境水中の食品添加物青色1号の定量
- LC-MSを用いるフェニルヒ素化合物の分析条件の検討と茨城県神栖市の地下水試料への適用