Determination of Brain Catecholamines by High Performance Liquid Chromatography with mewly developed Electrochemical Detector
スポンサーリンク
概要
- 論文の詳細を見る
This paper confirms the usefulness of electrochemical detector (ED) for liquid chromatography in determining picogram amounts of catecholamines in mouse and rat brain. The extraction procedure was based on modifications of Kissinger's method and of Anton and Sayre, in which catecholamines were adsorbed on aluminum oxide and eluted with perchloric acid. Thedetector operates down to the 0.2-0.4ng range. Norepinephrine (NE) Ievels were 431±18ng/g in mouse brain and 426±19ng/g in the rat. Dopamine (DM) levels were 1,088±66ng/g in mouse brain and 787±33ng/g in the rat. The authors have-fouad significant changes in DM levels in rat whole brain depending on the method of sacrifice. Levels of NE in the forebrain after electrolytic destruction of the locus coeruleus were significantly decreased (24.2%) from that of control.
- Japan Society of Clinical Chemistryの論文
著者
関連論文
- 超短波照射法による脳内酵素の失活法について
- 酵素アイソトープ法(radioenzymatic assay method)によるCatecholamine(CA)測定法の進展とその応用
- GC及びGC/MSによる米糖中ステロイド類似物質の同定ならびに定量